This note discusses sources of error in the use of charged-particle analyzers to determine the ionic species and energies in plasma streams, with particular reference to the widely used electrostatic instrument. Space-charge expansion within the analyzer leads to an upper limit on the size of the entrance aperture. This limit, which depends on the ion energy and density, is less than the proper Debye shielding distance that is often considered a sufficient limit. For ion energies below a few hundred eV and densities above 1010 cm−3 the aperture requirement is impractical. Therefore the possibility of reducing the plasma density by means of single apertures or multi-hole screens is considered. These considerations are supported by measurements on plasma gun outputs. The best criteria for proper operation of the analyzer are that the integral over the output spectrum equal the number of incident ions and that the output intensity be proportional to the area of the entrance hole.
本稿では、プラズマ流中のイオン種とエネルギーを決定するための荷電粒子アナライザーの使用における誤差の原因について論じ、特に広く用いられている静電型装置に言及する。アナライザー内の空間電荷膨張は、入口アパーチャのサイズに上限を設ける。この上限は、イオンエネルギーと密度に依存し、しばしば十分な限界と見なされる適切なデバイ遮蔽距離よりも小さい。イオンエネルギーが数百eV未満で密度が10¹⁰ cm⁻³を超える場合、アパーチャ要件は非現実的となる。そこで、単一アパーチャまたは多孔スクリーンによってプラズマ密度を低減する可能性を検討する。これらの考察は、プラズマガン出力の測定によって裏付けられている。アナライザーの適切な動作のための最良の基準は、出力スペクトル上の積分が入射イオン数に等しいこと、および出力強度が入口穴の面積に比例することである。