AbstractA vacuum crystal spectrometer that covers a wavelength range of 1.0–2.5 nm has been developed to observe light (oxygen) and heavy (titanium and iron) impurities for large helical device (LHD) diagnostics. A Johann-type spectrometer (2R = 60 cm and 3 m, RAP crystal, 2d = 2.612 nm) was constructed and installed in a compact helical system (CHS). The O VII resonance line was successfully observed through a thin (3.3 μmt) polypropylene film, although the metallic impurity lines were not seen. In the case of a crystal with 2R = 3 m, however, the spectral resolution λ/Δλ was limited to 440, whereas the replacement by the crystal with 2R = 60 cm improved the resolution to 610. The result is discussed with regard to the quality of the bent crystal.
大型ヘリカル装置(LHD)の診断のために、波長範囲1.0~2.5 nmをカバーする真空結晶分光器を開発し、酸素およびチタン、鉄などの重不純物の観測を行った。ヨハン型分光器(2R = 60 cmおよび3 m、RAP結晶、2d = 2.612 nm)をコンパクトヘリカルシステム(CHS)に設置した。O VII共鳴線は、厚さ3.3 μmのポリプロピレンフィルムを通して観測に成功したが、金属不純物の輝線は観測されなかった。しかし、2R = 3 mの結晶の場合、分光分解能λ/Δλは440に制限されたのに対し、2R = 60 cmの結晶に交換することで分解能は610に向上した。この結果について、湾曲結晶の品質の観点から考察を行った。