The analysis of hydrogen isotope content is crucial for understanding the operation of fusion devices. Hydrogen isotopic analysis in the core and plasma edge is conducted through neutron and spectroscopic diagnostics. In the case of exhaust gas, mass spectrometry is employed using residual gas analyzers (RGAs), along with optical gas analysis (OGA) utilizing an optical Penning gauge. The use of traditional quadrupole mass spectrometers for RGA encounters challenges during discharges with tritium gas due to signal overlap of different hydrogen molecules at the same mass number. This paper introduces a novel technique to address this issue through a cross-related analysis of RGA data with OGA results. Another consideration in mass spectrometry analysis is the instrument's varying sensitivity concerning the gas mass number. The paper includes gas calibration data results for all the quadrupoles used in the JET gas analysis. Hydrogen isotopic ratios are calculated from RGA detected currents using simple formulas. The results of this procedure are presented for selected DT JET discharges in relation to the JET pulse time, and they are compared with corresponding optical data. Time-averaged hydrogen isotopic ratio values are computed for numerous discharges during the DT and T campaigns.
水素同位体含有量の分析は、核融合装置の運転を理解するために極めて重要である。コアおよびプラズマ端部における水素同位体分析は、中性子および分光診断によって実施される。排気ガスの場合には、光学ペニングゲージを利用した光学ガス分析(OGA)とともに、残留ガス分析器(RGA)を用いた質量分析が採用される。RGAに従来の四重極質量分析計を使用すると、トリチウムガスを用いた放電中に、異なる水素分子が同じ質量数で信号が重なるという課題が生じる。本論文では、RGAデータとOGA結果の相互関連分析を通じてこの問題に対処する新し手法を紹介する。質量分析におけるもう一つの考慮事項は、ガスの質量数に関する装置の感度の変動である。本論文には、JETガス分析で使用されたすべての四重極のガス校正データ結果が含まれている。水素同位体比は、RGAで検出された電流から簡単な式を用いて計算される。この手順の結果は、選択されたDT JET放電についてJETパルス時間との関係で示され、対応する光学データと比較される。時間平均された水素同位体比の値は、DTおよびTキャンペーン中の多数の放電に対して計算される。