Single photon counting using a charge-coupled device (CCD) is a conventional method of measuring the x-ray production of laser plasma. But the spectrum is seriously distorted when measuring a short pulse laser plasma x-ray source. Here we explore this abnormal spectral distortion using a silicon sensor-based CCD (PIXIS-XB: 1300R). We found that the spectral distortion is caused by the continuous spectrum of Compton scattering electrons and the K-α and K-β characteristic lines of Ag elements excited by the harder x-rays of >25 keV. We calibrated the detection efficiency of the incident x-ray, the excitation efficiency of the Ag characteristic lines and the Compton scattering electrons spectrum in the x-ray energy range from 20 keV to 100 keV. We also improve the algorithm of spectrum reconstruction to loosen the exposure level limit markedly. Our effort extends the detecting range of silicon sensor-based CCDs from 0.5–30 keV to 0.5–80 keV. The present analysis can also be used in the spectral extraction of GaAs and CdTe high-Z sensor-based CCDs.
単一光子計数を用いた電荷結合素子(CCD)は、レーザープラズマのX線生成を測定する従来の方法である。しかし、短パルスレーザープラズマX線源を測定する際、スペクトルは著しく歪む。ここでは、シリコンセンサー型CCD(PIXIS-XB: 1300)を用いて、この異常なスペクトル歪みを調査する。我々は、このスペクトル歪みが、>25 keVの硬X線によって励起されたAg元素のK-αおよびK-β特性線と、コンプトン散乱電子の連続スペクトルによって引き起こされることを見出した。我々は、20 keVから100 keVのX線エネルギー範囲において、入射X線の検出効率、Ag特性線の励起効率、およびコンプトン散乱電子スペクトルを較正した。また、スペクトル再構成のアルゴリズムを改良し、露光レベルの制限を大幅に緩和した。我々の取り組みにより、シリコンセンサー型CCDの検出範囲は0.5–30 keVから0.5–80 keVへと拡張される。本解析は、GaAsおよびCdTeの高Zセンサー型CCDのスペクトル抽出にも適用可能である。