AbstractA new type of bolometric array diagnostic system has been developed for HT-6M tokamak and HT-7 superconductor tokamak plasma physics experimental research. This system is composed of temperature-sensitive detectors of gold film resistor and phase-sensitive rectifiers of the bridge excited by square waves. With this system, a radiation detection limit of 192.0 μW cm−2 at a spatial resolution of 2.0 cm and a temporal resolution of 1.0 mS. The system features a high temperature baking resistance, ultra high vacuum endurance, sufficient insensitivity to radiation damage, sufficient suppression of electromagnetic interference, good long-term stability, high radiation sensitivity and measurement data reliability. Absolute calibration of the detectors is performed simultaneously and in situ by means of a built-in electronic power pulse.
HT-6MトカマクおよびHT-7超伝導トカマクのプラズマ物理実験のために、新型ボロメトリックアレイ診断システムが開発された。本システムは、金薄膜抵抗の温度感応検出器と、矩形波によって励起されるブリッジの位相感応整流器から構成される。本システムにより、空間分解能2.0cm、時間分解能1.0mSにおいて、放射検出限界192.0μW cm⁻²が達成される。本システムは、高温ベーキング耐性、超高真空耐久性、放射損傷に対する十分な非感受性、電磁干渉の十分な抑制、良好な長期安定性、高い放射感度および測定データの信頼性を特徴とする。検出器の絶対較正は、内蔵電子パワーパルスにより同時かつその場で実施される。