AbstractA thin metal foil bolometric detector for time-resolved radiation loss measurements in the Heliotron E toroidal plasma confining experiment is described. The bolometer consists of a 10 or 50 gm thick stainless-steel radiation absorber, an insulation layer and a nickel resistor pattern. The bolometric detector, its electronics, the construction of the array and installation under vacuum and high-temperature conditions in toroidal plasma devices are reported. The physical principles of detection of the wavelength integrated total radiative loss from plasmas are described. The detector array consists of wide-angle bolometers to monitor the time-resolved total radiative loss power and a recently tested 17-channel bolometer array camera to measure the profiles of radiative loss. Typical radiative loss measurements in helical plasmas on Heliotron E with ECRH and NBI heating are presented.
ヘリオトロンEトーラス型プラズマ閉じ込め実験における時間分解放射損失測定用の薄膜金属ボロメータ検出器について述べる。ボロメータは、10または50μm厚のステンレス鋼放射吸収体、絶縁層、およびニッケル抵抗パターンから構成される。ボロメータ検出器、その電子回路、アレイの構造、ならびにトーラス型プラズマ装置における真空および高温条件下での設置について報告する。プラズマからの波長積分全放射損失の検出の物理的原理について述べる。検出器アレイは、時間分解全放射損失パワーを監視するための広角ボロメータと、放射損失の分布を測定するための最近試験された17チャンネルボロメータアレイカメラから構成される。ECRHおよびNBI加熱を伴うヘリオトロンEのヘリカルプラズマにおける典型的な放射損失測定結果を示す。